Features
- 12.1" High-resolution LED display with 8 waveforms
- Lightweight, less than 4.5 kg
- Easy to use for intuitive operation
- Plug-and-play modular design, more choices for clinical use
- No-fan design for a quiet care environment
- Trust for veterinary use, parameters including ECG, RESP, SpO2, NIBP, PR, TEMP, IBP, CO2, Multi-gas
- 360-degree visible alarm indicator
- 120 hours trend and 48 hours waveforms reviewing
Standard Configuration
- ECG
- RESP
- NIBP
- Mindray SpO2
- TEMP
- Li-ion battery
Specifications
- Weight: ≤4.5 kg (12 lb)
- Size: <280 x 320 x 130 mm (11 x 12.6 x 5.1 inch)
- Display: 12.1" high-resolution LED with 800 x 600 pixels
- Waveforms:Â Up to 8 traces
- ECG:Â 5 or 3 lead: I, II, III, aVR, aVL, aVF, V
- Respiration:Â RA-LA/RA-LL Impedance
- SpO2:Â Masimo SET, Nellcor Oximax
- NIBP:Â Manual, Auto, STAT
- Temperature:Â 2 channels
- EtCO2: Sidestream CO2 or Mainstream CO2
- Multi-gas option:Â Artema Gas CO2, O2 (option), N2O, Des, Iso, Enf, No Hal, Sev
- Recorder:Â 3-trace thermal array, 50 mm
- Reviewing:Â
- 120 hours trends
- 100 alarm events and associated waveforms
- 1000 groups NIBP
- Up to 48 hours waveforms
- Download to PC: USB
- Video Output: 1VGA
- Wired Network:Â 1 RJ45
ECG
- RR Range: 15 - 350 bpm
- Speed: 6.25/12.5/25/50 mm/s
- ECG Gain: 1.25/2.5/5/10/20/40 mm/mV, Auto Sweep
- Mode: Diagnostic, Monitor, or Surgical
PR Range
- 25 to 240 bpm (Masimo SpO2)
- 20 to 300 bpm (Nellcor SpO2)
- 25 to 350 bpm (IBP)
- 40 to 240 bpm (NIBP)
NIBPÂ Range
- Systolic 40 to 270 mmHg
- Diastolic 10 to 210 mmHg
- Mean 20 to 230 mmHg
- Interval in Auto Mode:
- 1/2/2.5/3/5/10/15/20/30/60/90/120/180/240/ 480min
- Max Time Allowed: 180 s
Respiration
- Lead: I, II or Auto
- RR Range: 0 to 150 rpm
- Accuracy: 7 to 150 rpm: ±2 rpm or ±2%, whichever is greater
Nellcor Oximax SpO2
- Measurement Range: 20-300 bpm
- Accuracy:
- 20 to 250 bpm: ±3 bpm
- 251 to 30 bpm: not specified
Masimo SET SpO2
- Measurement Range: 25-240 bpm
- Accuracy:
- ±3 bpm (measured without motion)
- ±5 bpm (measured without motion)
C.O.
- Range:
- C.O.: 0.1 to 20 L/min
- TB: 23 to 43°C
- TI: 0 to 27°C
- Accuracy:
- C.O.: ±5% or ±0.1 L /min whichever is greater
- TB, TI: ±0.1°C (without sensor)
- Resolution:
- C.O.: 0.1 L/min
- TB, TI: 0.1°C
IBP
- Range: -50 to 300 mmHg
- Accuracy: ±2% or±1mmHg whichever is greater
EtCO2 - Sidestream
- Sample Rate: 70 /100 ml/min
- Range: 0 to 99 mmHg
- awRR Range: 0 to 120 rpm
- Accuracy:
- 0 to 40 mmHg, ±2 mmHg
- 41 to 76 mmHg, ±5% of the reading
- 77 to 99 mmHg, ±10% of the reading
EtCO2 - Mainstream
- Range: 0 to 150 mmHg
- awRR Range: 0 to 150 rpm
- Accuracy:
- 0 to 40 mmHg, ±2 mmHg
- 41 to 70 mmHg, ±5% of the reading
- 101 to 150 mmHg, ±10% of the reading
- 71 to 100 mmHg, ±8% of the reading